Функциональный генератор TTi, TG2000
Aim-TTi TG1000 & TG2000 DDS Function Generators, 10MHz & 20MHz
The TG1000/2000 offers a high quality direct digital synthesis (DDS) function generator at a significantly lower price. DDS generates waveforms using a phase accumulator, look-up table and DAC. The accuracy/stability of the resulting waveform is related to that of the crystal master clock. When correctly engineered, the DDS generator offers not only exceptional accuracy and stability but also high spectral purity, low phase noise and excellent frequency agility. The TG2000 includes RS-232 and USB interfaces for remote control of all instrument functions and for remotely storing instrument set-ups.
6 digits or 1mHz setting resolution
1ppm stability and better than 10ppm absolute accuracy for one year
Sine, square, triangle, positive pulse and negative pulse waveforms
Internal sweep, linear or logarithmic, phase continuous, 0.1Hz to 20MHz
Modulations modes of sweep, gated, AM, FSK and tone switching - built-in trigger generator
5mV to 20V pk-pk output from 50Ω or 600Ω; plus multi-function auxiliary output
Storage for up to nine complete instrument set-ups in non-volatile memory
Sine Wave
Frequency Range 1mHz to 10MHz/20MHz; resolution 6 digits or 1mHz
Distortion <0.3 % THD to 500kHz, <-50dBc to 1MHz, <-30dBc to 20MHz
Square Wave and Positive/Negative Pulse
Frequency Range 1mHz to 10MHz/20MHz; resolution 6 digits or 1mHz
Rise & Fall Times < 22ns
Symmetry Variable 20 % to 80 % in 1 % steps
Triangle Wave
Frequency Range
Linearity Error
< 0.5 % to 30 KHz 1mHz to 1MHz; resolution 6 digits or 1mHz
Output Level
Sine/Square/Triangle 5mV to 20V pk-pk from 50Ω/600Ω source (switchable)
Pulse 2.5mV to 10V pk-pk from 50Ω/600Ω source, positive or negative pulses with respect to DC offset baseline
Accuracy & Flatness ± 3% ± 1mV at 1kHz into 50Ω/600Ω.
Flatness ± 0.2dB to 500kHz; ± 1dB to 20MHz
DC Offset ± 2.5mV to ± 10V from 50Ω/600Ω source
Internal Sweep
Sweep Width 0.1Hz to 20MHz; single range; lin or log; phase continuous
Sweep Time 20ms to 999s (3 digit resolution); triggered or continuous
Trigger and Marker Trigger from front panel or remote interface; variable marker output
Modulations
Modulation Modes External AM; Int/Ext Gate; Int/Ext FSK; Int/Ext Tone
Trigger Generation Trigger signals can be generated from an internal trigger generator, a front panel key, an external input, or a remote interface
Outputs and Inputs
Outputs Main - 50Ω/600Ω. switchable, Aux - Multifunction logic level output
Inputs VCA In - amplitude modulation I/P; External Trigger Gate I/P
Interfaces
RS-232 Variable baud rate, 19.200 max., 9 pin D connector
USB Conforming with USB1.1
General
Power Requirement 220-240V, 110-120V or 110V ±10 % 50/60Hz;
Installation Category II
Size and Weight 260 x 88 x 235mm. 2.0kg
Standards
Complies with EN61010-1 and EN61326
Supplied with
Mains lead and full instructions |